Human sTfR Quantikine IVD ELISA Kit Summary

深圳子科生物優(yōu)勢(shì)代理美國(guó)R&DELISA系列產(chǎn)品試劑盒,價(jià)格優(yōu)勢(shì),貨期短,歡迎來(lái)電咨詢(xún)0755-33164177!

貨號(hào):DTFR1

Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
2.5 hours
Sample Type & Volume Required Per Well
Serum (20 uL), EDTA Plasma (20 uL), Heparin Plasma (20 uL), Citrate Plasma (20 uL)
Sensitivity
0.5 nmol/L
Assay Range
3.0 - 80 nmol/L (Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma)
Specificity
Measures human sTfR levels in various samples
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
Interference observed with other substances.

Product Summary

The Quantikine IVD Human sTfR Immunoassay is designed for the quantitative determination of soluble transferrin receptor (sTfR) concentrations in human serum and plasma as an aid in the diagnosis of Iron-Deficiency Anemia (IDA), especially for the differential diagnosis of IDA and Anemia of Chronic Disease (ACD). For In Vitro Diagnostic Use.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision

Serum, EDTA Plasma, Heparin Plasma, Citrate Plasma

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 60 60 60
Mean 6.9 18.2 40.2 10.9 25.7 67.2
Standard Deviation 0.3 1.3 2.5 0.7 1.4 3.8
CV% 4.3 7.1 6.2 6.4 5.4 5.7

Recovery

The recovery of sTfR spiked to three different levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Citrate Plasma (n=5) 112 105-121
EDTA Plasma (n=5) 103 97-112
Heparin Plasma (n=5) 106 101-112
Serum (n=5) 97 95-100

Linearity

To assess the linearity of the assay, five samples containing high concentrations of sTfR were diluted with Specimen Diluent and evaluated.
Human TfR (Transferrin R) In Vitro Diagnostic ELISA Linearity